R&S AFQ100A
AFQ Commands - Setup Section
1401.3084.32
6.73
E-3
CALibration:AOUTput<CH>:DELay?
The command adjusts the amplitude of the (selected) analog output.
The instrument generates a series of signal levels and converts them using a precise ADC. The
procedure takes about 30 seconds.
The command is a query command and therefore has no *RST value.
Example:
CAL:AOUT:AMPL?
'starts all internal adjustments for which no external measuring equipment is
needed. The adjustments with external measuring equipment are described
in
the
service manual
.
Response: '0'
'calibration was successful
*RST value
Resolution
Options
SCPI
-
-
Device-specific
Diagnostic Subsystem
The DIAGnostic system contains the commands used for instrument diagnosis and servicing.
In the Test Point section provides access to the test points available in the instrument except for the
analog output board (see below). When activated, the voltage of the selected test point is measured
and displayed. A detailed description of the test points can be found in chapter 3 of the
service manual
.
In the Analog Output Board Test Board section provides access to the test points on the analog output
board
In the Fan Control section the fan speed is selected.
Command
Parameters
Default
unit
Remark
DIAGnostic:AOUTput<CH>:POINt:CATalog?
Query
only
DIAGnostic:AOUTput<CH>:POINt?
Query only
DIAGnostic:BGINfo?
<module name>
Query only
DIAGnostic:FCONtrol
AUTO, FSPeed
DIAGnostic:INFO:POCounter?
Query only
DIAGnostic:INFO:OTIMe?
Query only
DIAGnostic[:MEASure]:POINt?
Query only
DIAGnostic:POINt:CATalog?
Query only