Remote control commands
R&S
®
NRPxxP
51
User Manual 1179.5760.02 ─ 02
The delay time is set using
Parameters:
<state>
*RST:
OFF
TRIGger:COUNt
<count>
Sets the number of measurement cycles to be performed when the measurement is
started using
.
This number equals the number of results that can be obtained from the power sensor
after a single measurement. As long as the defined number of measurements is not
executed, the power sensor automatically initiates another measurement internally
when the current result is available.
This command is particularly useful in conjunction with buffered measurements. For
example, to fill a buffer with a predefined size with measurements that have been trig-
gered externally or by
without having to start the measurement multiple times.
Parameters:
<count>
Range:
1 to 2147483646
*RST:
1
Example:
See
Chapter 9.2, "Performing a buffered continuous average
TRIGger:DELay
<delay>
Sets the delay between the trigger event and the beginning of the actual measurement
(integration).
Parameters:
<delay>
Range:
-5.0 to 10.0
*RST:
0.0
Default unit: s
TRIGger:DTIMe
<dropout_time>
Sets the dropout time for the internal trigger source. During this time, the signal power
must exceed (negative trigger slope) or undercut (positive trigger slope) the level
defined by the trigger level and trigger hysteresis. At least, this time must elapse before
triggering can occur again.
See
Chapter 8.5.2.4, "Dropout time"
Parameters:
<dropout_time>
Range:
0.00 to 10.00
*RST:
25.0e-9
Default unit: s
Controlling the measurement