Command reference
R&S
®
ZNA
918
User Manual 1178.6462.02 ─ 20
Parameters:
<Boolean>
*RST:
OFF
Options:
R&S
ZNA-K20
Manual operation:
See
CALCulate<Chn>:EYE:MASK:VIOLation:RATE
<ViolationRate>
Defines the violation rate (i.e. the share of bad samples) for the mask test in the related
eye diagram.
This command will raise an execution error if the active trace in the selected channel is
not an eye diagram.
This rate will only be used if
CALCulate<Chn>:EYE:MASK:FAIL:CONDition
to RATE.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<ViolationRate>
Default unit: percent
Options:
R&S
ZNA-K20
Manual operation:
See
CALCulate<Chn>:EYE:MASK:VIOLation:TOLerance
<ViolationTolerance>
Defines the violation tolerance (i.e. the number of bad samples) for the mask test in the
related eye diagram.
This command will raise an execution error if the active trace in the selected channel is
not an eye diagram.
This tolerance will only be used if
CALCulate<Chn>:EYE:MASK:FAIL:CONDition
is set to SAMPles.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<ViolationTolerance>
Options:
R&S
ZNA-K20
Manual operation:
See
CALCulate<Chn>:EYE:MEASurement:DATA?
Returns the measurement results of the related eye diagram (see
SCPI command reference