GUI reference
R&S
®
ZNA
823
User Manual 1178.6462.02 ─ 20
Log Errors
Enables the transfer of error messages for external devices (e.g. connection errors) to
the error log. The error log appears in the "Info" dialog; see
Remote command:
n/a
5.19.5.2
External Generators dialog
The "External Generators" dialog configures external generators so that they can be
used for measurements and power calibrations.
Access:
System – [Setup] > "Generator" > "Generators..."
Background information
Chapter 4.7.39, "External generators"
The configuration of a new external generator involves the following steps:
1. Connect the generator to your R&S
ZNA using a LAN (VXI-11), GPIB, or USB inter-
face.
2. If the generator is connected via LAN, enable "LAN detection"
3. Click "Scan Instruments" and wait until the generator appears in the table of
"Known Devices".
4. Click to copy the generator into the list of configured devices.
Setup softtool