GUI reference
R&S
®
ZNA
791
User Manual 1178.6462.02 ─ 20
Interpolation is used if the sweep points during the measurement differ from the sweep
points during calibration (state label "Cal int"). Compared to the default linear interpola-
tion, the "Higher Order" (cubic spline) interpolation can slightly slow down the mea-
surement but can provide better results, at least if the measured trace shows no signifi-
cant noise or discontinuities. For the factory system error correction, the analyzer
always uses linear interpolation.
Generally, it is recommended to avoid interpolation whenever possible.
Remote command:
User Interface tab
Provides general user interface configurations.
Sounds
Switches the sounds generated by the analyzer GUI on or off. Sounds are generated
for certain GUI events, when the analyzer generates a notice, status or warning mes-
sage (alarm sounds) or during calibration.
Limit fail beeps for
,
, and
is not affected by this setting.
If enabled, those beeps are generated even if "Sounds" is unchecked.
Note:
The R&S
ZNA does not have a built-in audio device and loudspeaker. To hear
these sounds, connect a USB audio device to the instrument or operate it via remote
desktop.
Remote command:
SYSTem:SOUNd:ALARm[:STATe]
SYSTem:SOUNd:STATus[:STATe]