GUI reference
R&S
®
ZNA
589
User Manual 1178.6462.02 ─ 20
Note:
The exported file can be directly unzipped to the METAS VNA Tools user data
folder (default:
C:\Users\Public\Documents\Metas.Vna.Tools
). After
(re-)opening the METAS VNA Tools, the exported project and its associated static data
are available for offline analysis.
METAS Reconnection
.
This button is only enabled, if a METAS cal is active.
5.11.5.2
METAS Calibration (Manual) dialog
This dialog allows you to set up and perform a
nel.
Access:
Channel – [Cal] > "METAS Cal" > METAS Calibration – "Start (Manual)"
Only a single VNA channel can be METAS calibrated at a time. If you have a METAS
calibration for another channel (no matter in which recall set), this METAS calibration is
deleted when you apply the new one.
Step 1: Ports and Type
In step 1, choose the ports to be calibrated and the calibration type to be performed.
Only calibration types supported by METAS VNA Tools are available.
Ports
Selects the test ports to be calibrated.
Note:
Calibration and port de-/activation.
Cal softtool