Command reference
R&S
®
ZNA
1451
User Manual 1178.6462.02 ─ 20
SOURce<Ch>:POWer<PhyPt>:CORRection:DATA:PARameter<Wv>:COUNt?
Gets the number of available power calibrations (both source power and measurement
receiver).
Use
SOURce<Ch>:POWer:CORRection:DATA:PARameter<Wv>?
to retrieve the
related settings.
Suffix:
<Ch>
.
Number of the calibrated channel.
<PhyPt>
This suffix is ignored.
<Wv>
This suffix is ignored.
Usage:
Query only
Manual operation:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:DEFault
<arg0>
Enables the analyzer to create a set of default source power calibration data. The ana-
lyzer uses the reference channel power to acquire the default calibration data. No
external power meter is required.
Tip:
The main purpose of the default calibration data set is to provide a dummy power cali-
bration which you can replace with your own, external power calibration data. You may
have acquired the external data in a previous session or even on an other instrument.
If you want to use the external power calibration data on the analyzer, generate the
default data set first and overwrite it with the external data. For details refer to the pro-
gram example below.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored, the chan-
nel settings are relevant only when the default data set is cre-
ated
(
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
).
<PhyPt>
Calibrated port number. This parameter is ignored.
Parameters:
<arg0>
Enables (ON) or disables (OFF) the dummy source power cali-
bration for port number
<PhyPt>
. OFF means that a real source
power calibration is performed, for which an external power
meter must be connected to the analyzer.
*RST:
OFF
SCPI command reference