Command reference
R&S
®
ZNA
1450
User Manual 1178.6462.02 ─ 20
*RST:
RRAFter
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
Manual operation:
See
"Source Leveling/Src. Lev. Method"
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:RRECeiver
<arg0>
Enables or disables calibration of the reference receiver together with the source
power calibration.
Note:
It is not possible to disable flatness calibration
(
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:FLATness
) and the refer-
ence receiver calibration simultaneously.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the verifi-
cation is either enabled or disabled for all channels.
<PhyPt>
Calibrated port number. This parameter is ignored; the verifica-
tion is either enabled or disabled for all sources.
Parameters:
<arg0>
Enables (ON) or disables (OFF) reference receiver calibration.
*RST:
ON
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:
SOURce:POWer:CORRection:COLLect:VNETworks
<DeEmbNwType>[,
<DeEmbType>]
Defines whether the source flatness cal uses the configured offset de-/embedding to
get the power values at the DUT (instead of the calibrated reference plane).
This setting applies to all physical ports and all channels.
Parameters:
<DeEmbNwType>
SENDed | GLOop | OFF | ALL | DIFFerential | PSET | RENorm |
OFFSet | AOFFset
Selects the offset or de-/embedding type to be used, if config-
ured.
ALL means that all available offset and de-/embedding informa-
tion is applied. The <DeEmbType> can be omitted in this case.
*RST:
OFF
<DeEmbType>
EMBedding | DEEMbedding
Selects between embedding and deembedding for the selected
<DeEmbNwType>.
Must be specified for network types that can be used for embed-
ding and deembedding
Manual operation:
See
SCPI command reference