Command reference
R&S
®
ZNA
1438
User Manual 1178.6462.02 ─ 20
Manual operation:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:FREQuency:STARt
<Leveling Tool Start Frequency>
SOURce:POWer<PhyPt>:CORRection:SLEVeling:FREQuency:STOP
<FreqValue>
Defines the start and stop frequency of the
grid for physical port <PhyPt>.
By default the vector network analyzer uses the standard frequency range of the
selected "Meas Port".
After defining the leveling grid (using
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:FREQuency
and
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:POWer
commands), you
can start the leveling data acquisition using
SOURce:POWer<PhyPt>:CORRection:
.
Suffix:
<PhyPt>
.
Physical port number
Parameters:
<FreqValue>
Start/stop value
Default unit: Hz
Example:
See
SOURce:POWer<PhyPt>:CORRection:SLEVeling[:
Options:
R&S
ZNA-K8
Manual operation:
See
"Start Frequency/Stop Frequeny"
SOURce:POWer<PhyPt>:CORRection:SLEVeling:POWer:ALC
<Compatible>
If checked, the power step size of the
grid for physical port <PhyPt> is fixed to
1
dB. The step size defined using
SOURce:POWer<PhyPt>:CORRection:
is ignored.
After defining the leveling grid (using
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:FREQuency
and
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:POWer
commands), you
can start the leveling data acquisition using
SOURce:POWer<PhyPt>:CORRection:
.
Suffix:
<PhyPt>
.
Physical port number
Parameters:
<Compatible>
Example:
See
SOURce:POWer<PhyPt>:CORRection:SLEVeling[:
Options:
R&S
ZNA-K8
Manual operation:
See
SCPI command reference