Command reference
R&S
®
ZNA
1433
User Manual 1178.6462.02 ─ 20
SOURce<Ch>:POWer<Pt>:ALC:PIParameter:ITIMe
<Alc control loop coefficients
itime>
Defines the integration time of the ALC PI controller at source port <Pt> in channel
<Ch>.
This setting only takes effect if
SOURce<Ch>:POWer<Pt>:ALC:PIParameter
to
MANual
.
Suffix:
<Ch>
.
Channel number
<Pt>
Port number
Parameters:
<Value>
Integration time Ki [s]
Range:
1E-8 to 1E2
*RST:
0.4
Manual operation:
See
SOURce<Ch>:POWer<Pt>:ALC:RANGe
<CRange>
Defines the maximum change of the source signal level due to the ALC for physical
port <Pt> in channel <Ch>.
Suffix:
<Ch>
.
Channel number
<Pt>
Port number
Parameters:
<CRange>
Control range [dB]
Range:
0 to 100
Increment: 0.1
*RST:
5
Manual operation:
See
SOURce<Ch>:POWer<Pt>:ALC:RECeiver
<ALCReceiver>
Defines the receiver to be used for ALC at port <Pt> in channel <Ch>.
The target level is taken from the port configuration (
ON
.
Suffix:
<Ch>
.
Channel number
<Pt>
Port number
Parameters:
<ALCReceiver>
'A<n>'
Reference receiver at analyzer port <n>
SCPI command reference