Command reference
R&S
®
ZNA
1397
User Manual 1178.6462.02 ─ 20
[SENSe:]SWEep:COUNt:ALL
<Sweep>
Defines the number of sweeps to be measured in single sweep mode
(
OFF
). The setting is applied to all channels. Use
to define the sweep count for a single channel.
Setting parameters:
<Sweep>
Number of consecutive sweeps to be measured.
Range:
1 to 100000
*RST:
1
Example:
See
Chapter 8.2.4, "Data handling"
Usage:
Setting only
Manual operation:
See
[SENSe<Ch>:]SWEep:AXIS:FREQuency
<Scale>
Selects either the channel base frequency or one of the port frequencies as the stimu-
lus axis in all diagrams of the active recall set. This command has no effect if no port-
specific frequencies are defined.
Suffix:
<Ch>
.
Channel number
Parameters:
<Scale>
String parameter, one of the following signals:
'Channel Base; Source' - base channel frequency
'Port 1; Source' - source frequency at port 1
'Port 2; Source' - source frequency at port 2
...
'Gen 1; Source' - external generator frequency 1
...
'Pmtr 1; Receiver' - external power meter frequency 1
...
*RST:
'Channel Base: Source'
SCPI command reference