Command reference
R&S
®
ZNA
1384
User Manual 1178.6462.02 ─ 20
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STARt
<StartFreq>
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STOP
<StopFreq>
These commands define the start and stop frequency of sweep segment no.
<Seg>
.
The sweep segments must be within the frequency range of the R&S
ZNA; see
.
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number
Parameters:
<StopFreq>
Start or stop frequency of the sweep.
Increment: 0.1 kHz
Default unit: Hz
Example:
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0,
10KHZ
Create a sweep segment with a sweep range between 1.0 MHz
and 1.5 MHz.
SEGM:FREQ:STAR?; STOP?
Query the start and stop frequency of the created segment. The
response is
1000000;1500000
.
Manual operation:
See
Note:
If the start frequency entered is greater than the current stop frequency, the stop
frequency is set to the start frequency. If the stop frequency entered is smaller than the
current start frequency, the start frequency is set to the stop frequency.
[SENSe<Ch>:]SEGMent<Seg>:INSert
<StartFreq>, <StopFreq>, <Points>, <Power>,
<SegmentTime>|<MeasDelay>, <Unused>, <MeasBandwidth>[, <LO>,
<Selectivity>, <Mode>, <FreqSweepMode>]
Adds a new sweep segment no.
<Seg>
with specific channel settings.
Entry of the first seven numeric parameters is mandatory; no default values are provi-
ded. All settings except
<LO>
can be changed for existing segments using other com-
mands of the
[SENSe<Ch>:]SEGMent<Seg>...
subsystem.
Note:
to create a segment with default
channel settings. Use
[SENSe<Ch>:]SEGMent<Seg>:DEFine
settings of an existing segment.
Suffix:
<Ch>
.
Channel number
SCPI command reference