Command reference
R&S
®
ZNA
1354
User Manual 1178.6462.02 ─ 20
ON (1)
Manual selection using
Options:
R&S
ZNA-K30
Manual operation:
See
[SENSe<Ch>:]NFIGure:NTIMe
<DetectorTime>
Defines the detector time of the RMS detector which is used for the actual noise mea-
surement during a NF measurement in channel <Ch>
Suffix:
<Ch>
.
Channel number
Parameters:
<DetectorTime>
Detector time for the RMS detector
Default unit: s
Example:
SENSe1:NFIGure:NTIMe 50ms
Options:
R&S
ZNA-K30
Manual operation:
See
[SENSe<Ch>:]NFIGure:RFICorr
<Boolean>
Activates or deactivates the image band correction at the DUT, during a frequency con-
verting noise figure measurement in Channel <Ch>.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean>
ON (1)
image band correction activated
OFF (0)
image band correction deactivated
*RST:
OFF (0)
Example:
SENSe1:NFIGure:RFICorr ON
Options:
R&S
ZNA-K30
Manual operation:
See
[SENSe<Ch>:]NFIGure:TEATtenuator
<Temperature>
Sets the ambient temperature used for proper determination of the DUT input source
noise during a noise figure measurement in channel <Ch>.
Suffix:
<Ch>
.
Channel number
SCPI command reference