Command reference
R&S
®
ZNA
1110
User Manual 1178.6462.02 ─ 20
Example:
*RST; :CONF:CHAN2:STAT ON; NAME 'New Channel'
Create channel 2 and assign the channel name "New Channel".
CONF:CHAN:CAT?
Query all channels and their names. As a default channel no. 1
is created on
*RST
, the response is
'1,Ch1,2,New_Channel'
.
CONF:CHAN:NAME:ID? 'New Channel'
Query the channel number for the channel named "New Chan-
nel". The response is
2
.
Usage:
Query only
Manual operation:
See
CONFigure:CHANnel:MEASure:ALL[:STATe]
<Boolean>
Enables or disables the sweep in all channels of the active recall set. This command
can be used in combination with
CONFigure:CHANnel<Ch>:MEASure[:STATe]
optimize the measurement speed.
Parameters:
<Boolean>
ON | OFF
*RST:
ON
Example:
See
CONFigure:CHANnel<Ch>:MEASure[:STATe]
Manual operation:
See
CONFigure:CHANnel:MEASure:OPTimized
<Order>[, <Ch1>, <Ch2>, ...]
This setting tells the firmware to optimize the channel switching times for a particular
channel sweep order.
Note
that the optimization can only be applied, if
all
channels in the current setup are
in single sweep mode (
sweep sequence. Any change in the current setup or a wrong sweep sequence lets the
analyzer fall back to unoptimized channel switching.
Setting parameters:
<Order>
AUTO | MANual
AUTO
Optimize the switching times for the channel order that is auto-
matically used with
numbers).
MANual
Optimize the switching times for the channel order manually
specified using channel numbers <Ch1>, <Ch2>, ...
SCPI command reference