RF connections
R&S
®
ZCxxx
20
Getting Started 1177.5156.02 ─ 05
5
RF connections
The R&S
ZNA offers various possibilities to connect and operate one or more fre-
quency converters R&S
ZCxxx.
In the following descriptions, we focus on the
standard setups
for 2 and 4 con-
verters, which use the following hardware options:
●
Direct IF access (R&S
ZNA-B26)
●
Converter LO output (R&S
ZNA-B8)
With the standard setup, two (four) converters can be operated with a two-port
(four-port) R&S
ZNA, without external generator.
Reference & measurement
signals to IF rear inputs
(cables included in delivery)
LO from
R&S ZNA-B8
RF from test ports
R&S ZNA-B8 (rear panel LO output)
R&S ZNA-B26 (Direct IF I/O)
R&S ZNA-K8
R&S Z
V
-Z193
R&S ZV-Z193
R&S ZV-Z193
R&S ZV-Z195
Var. 37
Port adapters
required for R&S ZNA50/67
(included in R&S ZCAKN)
DC power supply
R&S ZCPS
(cables
ZCPS to ZCxxx
included)
Splitter 1-to-2
(included in R&S ZCAKN)
R&S Z
V
-Z193
Port adapters
(included in R&S ZCAKN)
Figure 5-1: Standard 2-port setup (with Rohde & Schwarz accessories)