Command reference
R&S
®
ZNB/ZNBT
856
User Manual 1173.9163.02 ─ 62
Query parameters:
<Data>
ALL
If omitted, a single numeric value is returned. If ALL is specified,
the result consists of 6 numeric values. Furthermore, the inter-
pretation of the result values depends on the active trace's stim-
ulus axis (
CALCulate<Chn>:TRANsform:TIME:XAXis
). See
the table below.
Note
that the skew value is positive, if, at the defined position
(
the right of the reference trace.
Usage:
Query only
Options:
R&S
ZNB/ZNBT-K20
Manual operation:
See
<Data> omitted
<Data> = ALL
Stimulus axis:
time
<skew time>
<skew time>,<skew distance>, <X value of ref. point on current
trace>, <Y value of ref. point on current trace>, <X value of ref.
point on reference trace>, <Y value of ref. point on reference
trace>
Stimulus axis:
distance
<skew distance>
<skew distance>, <skew time>, <X value of ref. point on cur-
rent trace>, <Y value of ref. point on current trace>, <X value of
ref. point on reference trace>, <Y value of ref. point on refer-
ence trace>
CALCulate<Chn>:DTIMe:LIMit:FAIL?
<Boolean>
Indicates whether the skew has passed or failed.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Return values:
<Boolean>
0 – skew check has passed
1 – skew check has failed
Usage:
Query only
Options:
R&S
ZNB/ZNBT-K20
CALCulate<Chn>:DTIMe:LIMit:FAIL:BEEP
<Boolean>
Deprecated command
: Since version 3.45, generating sounds is no longer supported
by the analyzer firmware.
Defines whether the R&S
ZNB/ZNBT makes an audible beep on skew limit violations.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
*RST:
OFF
SCPI command reference