Command reference
R&S
®
ZNB/ZNBT
1484
User Manual 1173.9163.02 ─ 62
[SENSe<Ch>:]FREQuency:MODE
<FreqSweep>
Selects the sweep type and defines which set of commands controls the stimulus fre-
quency.
Tip:
The command
provides a complete list of sweep
types.
Suffix:
<Ch>
.
Channel number
Parameters:
<FreqSweep>
CW | FIXed | SWEep | SEGMent
Linear or logarithmic frequency sweep, depending on the
selected spacing (
LINear |
LOGarithmic
). The frequency range is set via
Example:
FREQ:MODE CW
Activate a time sweep.
FREQ:CW 100MHz
Set the CW frequency to 100 MHz.
[SENSe<Ch>:]SEGMent<Seg>:CLEar
Deletes all sweep segments in the channel. The command is equivalent to
[SENSe<Ch>:]SEGMent<Seg>:DELete:ALL
.
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number. This suffix is ignored; the command
deletes all segments.
Usage:
Event
[SENSe<Ch>:]SEGMent<Seg>:OVERlap
<Boolean>
Queries whether the analyzer supports overlapping sweep segments.
Suffix:
<Ch>
.
Channel number. This suffix is ignored; the command is instru-
ment-specific.
<Seg>
Sweep segment number. This suffix is ignored; the command is
instrument-specific.
Parameters:
<Boolean>
ON | OFF: No effect.
*RST:
ON. If used as a query, the command returns the
information that overlapping sweep segments are
supported (ON).
R&S ZVR/ZVABT compatible commands