Command reference
R&S
®
ZNB/ZNBT
1362
User Manual 1173.9163.02 ─ 62
Example:
*RST; :SWE:POIN 10
Reset the instrument and reduce the number of sweep points to
10.
SOUR:POW:CORR:ACQ PORT,1
Perform a source power calibration using port 1 as a source
port.
SOUR:POW:CORR:DATA? 'A1'
Query the correction values. The analyzer returns 10 comma-
separated real numbers.
SOUR:POW:CORR:DATA 'A1', 1, 2, 3, 4, 5, -6, -7,
-8, -9, -0
Replace the correction values by ten (new) numbers.
Manual operation:
See
SOURce<Ch>:POWer:CORRection:DATA:PARameter<Wv>?
<SettingType>[,
<Index>]
Gets settings of the available power calibrations.
Suffix:
<Ch>
.
Number of the calibrated channel.
<Wv>
Number of the power calibration.
Use
SOURce<Ch>:POWer<PhyPt>:CORRection:DATA:
on page 1363 to query the number
of available power calibrations.
Query parameters:
SettingType
WAVE | STARt | STOP | POINts | STYPe | ATTenuation |
CPOWer | CFRequency | TSTamp | LTSTamp | TVNA | MVNA |
MTESt | FSMode
WAVE
Calibrated port and calibration type (Gen=Source Power,
Rec=Measurement Receiver)
STARt
Start Frequency or Power
Stop Frequency or Power
POINts
Nr. of sweep points
STYPe
Sweep type or grid (LIN, LOG, SEGM)
ATTenuation
Source or receiver attenuation
CPOWer
Cal Power for frequency sweeps
CFRequency
Cal Frequency for power sweeps
SCPI command reference