Command reference
R&S
®
ZNB/ZNBT
1082
User Manual 1173.9163.02 ─ 62
Note:
*RST
or "Preset" do not change the configuration of the
to restore
default values.
*RST:
n/a (default: 0 (port A, B, and F); ports C, D, and E
are configured as input ports.)
Example:
CONT:HAND:A:MODE OUTP
Configure port A as an output port.
CONT:HAND:A 192
Write data to port A.
CONT:HAND:B:MODE INP
Configure port B as an input port.
CONT:HAND:B?
Read data from port B.
CONTrol:HANDler:A:MODE
<Mode>
CONTrol:HANDler:B:MODE
<Mode>
CONTrol:HANDler:C:MODE
<Mode>
CONTrol:HANDler:D:MODE
<Mode>
Controls the direction of the data flow at ports A, B, C, D. The direction at the com-
bined ports E, F, G, H is according to the configuration at the other ports.
Parameters:
<Mode>
INPut | OUTPut
INPut – Input of data at the port
OUTPut – Output of data at the port
Note:
*RST
or "Preset" do not change the configuration of the
to restore
default values.
*RST:
n/a (default: Port A and B: OUTPut (also valid for
port F); port C and D: INPut (also valid for port E).
Ports G and H have mixed default modes.)
Example:
See
CONTrol:HANDler[:EXTension]:INDex:STATe
<Boolean>
Selects the digital signal that is routed to pin 20 of the Universal Interface connector.
Parameters:
<Boolean>
ON - /INDEX signal at pin 20
OFF - /PORT_B6 signal at pin 20
*RST:
n/a (default: OFF)
Note:
*RST
or "Preset" do not change the configuration of the
to restore
default values.
SCPI command reference