R&S ESU
Remote Control – Programming Examples
More Complex Programming Examples
7.26
Operating Manual 1302.6163.12 - 03
7.3.8
Level Correction of Transducers (Definition of Transducer Fac-
tors)
In more complex test systems, the frequency response of the test setup must be
taken into account in all power measurements to avoid any measurement errors
being introduced from sources other than the DUT.
The R&S ESU offers the possibility of defining a frequency-dependent attenuation
correction factor (transducer factor).
In the example below, a factor with the following characteristics is defined:
Name: Transtest
Unit: dB
Scaling: lin
Comment: simulated cable correction
The factor is defined and can be activated as required.
REM ************************************************************************
Public Sub TransducerFactor()
'--------- Define transducer factor -------------------------
status = viWrite(analyzer, "CORR:TRAN:SEL 'TRANSTEST'", 25, retCnt)
'Define "Transtest" transducer factor
status = viWrite(analyzer, "CORR:TRAN:UNIT 'DB'", 19, retCnt)
'Unit 'dB'
status = viWrite(analyzer, "CORR:TRAN:SCAL LIN", 18, retCnt)
'Linear frequency axis
status = viWrite(analyzer, "CORR:TRAN:COMM 'Simulated cable correction'", 43,
retCnt)
cmd$ = "CORR:TRAN:DATA "
'Enter frequency and level
cmd$ = cmd$ + "10MHz, 0,"
'values. Level values without unit!
cmd$ = cmd$ + "100MHz, 3,"
cmd$ = cmd$ + "1GHz, 7,"
cmd$ = cmd$ + "3GHz, 10"
status = viWrite(analyzer, cmd$, 50, retCnt)
'Enter frequency and level values
'--------- Activate transducer ------------------------------
status = viWrite(analyzer, "CORR:TRAN:STAT ON", 17, retCnt)
'Activate transducer factor
END SUB
REM ************************************************************************
Frequency
Level
10 MHz
0 dB
100 MHz
3 dB
1 GHz
7 dB
3 GHz
10 dB
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...