Operating Manual 1302.6163.12 - 03
6.239
R&S ESU
Remote Control – Description of Commands
SYSTem Subsystem
Example:
"SYST:COMM:PRIN:ENUM:NEXT?"
Characteristics:
*RST value: NONE
SCPI: device-specific
Mode:
R, A
SYSTem:COMMunicate:PRINter:SELect <1|2>
<printer_name>
This command selects one of the printers configured under Windows NT including the associated
output destination.
The specified printer name must be a string as returned by the commands
SYSTem:COMMunicate:PRINter:ENUMerate:FIRSt?
or
SYSTem:COMMunicate:PRINter:ENUMerate:NEXT?
Example:
"SYST:COMM:PRIN:SEL 'LASER on LPT1'"
Characteristics:
*RST value: NONE
SCPI: device-specific
Mode:
R, A
SYSTem:DATE
1980 to 2099, 1 to 12, 1 to 31
This command is used to enter the date for the internal calendar.
The sequence of entry is year, month, day.
Example:
"SYST:DATE 2000,6,1"
Characteristics:
*RST value: –
SCPI: conform
Mode:
R, A
SYSTem:DISPlay:FPANel
ON | OFF
This command activates or deactivates the display of the front panel keys on the screen.
With the display activated, the instrument can be operated on the screen using the mouse by pressing
the corresponding buttons. This may be useful if the instrument is operated in a detached station by
means of a remote program such as PCANYWHERE.
Example:
"SYST:DISP:FPAN ON"
Command
HCOPy:DESTination
is used to select an output medium other than
the default one.
With the display of the front panel keys activated, the screen resolution of the unit
is set to 1024x768. Thus, only a section of the whole screen is visible on the inter-
nal LCD display, which will be moved by mouse moves.
For a full display of the user interface, an external monitor has to be connected to
the rear panel.
When the front panel display is deactivated, the original screen resolution is
restored.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...