Operating Manual 1302.6163.12 - 03
6.225
R&S ESU
Remote Control – Description of Commands
SOURce Subsystem
Mode:
R, A
SOURce<1|2>:EXTernal<1|2>:FREQuency:SWEep[:STATe]
ON | OFF
This command activates or deactivates the frequency sweep for generator 1 or 2 in the selected
window.
This command is only valid in combination with option Ext. Generator Control R&S FSP-B10.
Example:
"SOUR:EXT1:FREQ:SWE ON"
'Activates the frequency sweep for ext. generator 1.
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
R, A
SOURce<1|2>:EXTernal<1|2>:POWer[:LEVel]
<numeric_value>
This command sets the output power of the selected generator 1 or 2 in the selected window.
This command is only valid in combination with option Ext. Generator Control R&S FSP-B10.
Example:
"SOUR:EXT:POW –30dBm"
'Sets the generator level to –30 dBm
Characteristics:
*RST value: -20 dBm
SCPI: device-specific
Mode:
R, A
SOURce<1|2>:EXTernal<1|2>:ROSCillator[:SOURce]
INTernal | EXTernal
This command switches between external and internal reference oscillator for the frequency
processing of external generator 1 and 2.
The command always works on both generators. Therefore, the numeric suffix in EXTernal<1|2> is not
significant.
This command is only valid in combination with option Ext. Generator Control R&S FSP-B10.
Example:
"SOUR:EXT:ROSC EXT"
'switches to external reference oscillator
Characteristics:
*RST value: INT
SCPI: device-specific
Mode:
R, A
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...