Operating Manual 1302.6163.12 - 03
6.205
R&S ESU
Remote Control – Description of Commands
SENSe Subsystem
Example:
"POW:ACH:PRES:RLEV;*WAI"
'Adapts the reference level to the measured channel power.
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
A-F
[SENSe<1|2>:]POWer:BANDwidth|BWIDth
10 to 99.9PCT
This command defines the percentage of the power with respect to the total power. This value is the
basis for the occupied bandwidth measurement (command:
POWer:ACHannel:PRESet OBW
).
The command is only available in the frequency domain (span > 0).
Example:
"POW:BWID 95PCT"
Characteristics:
*RST value: 99PCT
SCPI: device-specific
Mode:
A-F
[SENSe<1|2>:]POWer:HSPeed
ON | OFF
This command switches on or off the high-speed channel/adjacent channel power measurement. The
measurement itself is performed in the time domain on the center frequencies of the individual
channels. The command automatically switches to the time domain and back.
Depending on the selected mobile radio standard, weighting filters with
√
cos characteristic or very
steep-sided channel filters are used for band limitation.
The command is only available in the frequency domain (span > 0).
Example:
"POW:HSP ON"
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A-F
Subsequent commands have to be synchronized with
*WAI
,
*OPC
or
*OPC?
to the
end of the auto range process which would otherwise be aborted.
If the high-speed power measurement is switched off, the complete channel/adja-
cent channel power measurement is switched off.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...