R&S ESU
Remote Control – Description of Commands
SENSe Subsystem
6.190
Operating Manual 1302.6163.12 - 03
[SENSe<1|2>:]LIST:RANGe<1…20>[:FREQuency]:STOP
<numeric_value>
This command sets the stop frequency of a range in the spurious measurement.
Example:
"LIST:RANG2:STOP 2GHZ"
'Stop frequency of range 2 to 2 GHz
Characteristics:
*RST value: --
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]LIST:RANGe<1…20>:INPut:ATTenuation
<numeric_value>
This command defines the RF attenuation of a range in the spurious measurement.
Example:
"LIST:RANG2:INP:ATT 30db"
'RF attenuation of range 2 to 30 dB
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]LIST:RANGe<1…20>:INPut:ATTenuation:AUTO
ON | OFF
This command switches the autoranging of a range in the spurious measurement on or off.
Example:
"LIST:RANG2:INP:ATT:AUTO ON"
'Activates autorange for range 2
Characteristics:
*RST value: ON
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]LIST:RANGe<1…20>:INPut:GAIN:STATe
ON | OFF
This command switches the preamplifier of a range in the spurious measurement on or off.
Example:
"LIST:RANG2:INP:GAIN:STAT ON"
'Activates the preamplifier for range 2
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]LIST:RANGe<1...20>:LIMit
-200dB to +200dB
This command defines the limit lines for the spurious emission measurement
(for
details on this measurement see page 220). For each of the 20 ranges, you can specify a limit line via
the numeric suffix at RANGe<1...20>. The numeric suffix at SENSe<1|2> is irrelevant.
Example:
"LIST:RANG5:LIM -40"
’Limit line in range 5 is –40 dB.
Characteristics:
*RST value: -13dB
SCPI: device-specific
Mode:
A
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...