Operating Manual 1302.6163.12 - 03
6.151
R&S ESU
Remote Control – Description of Commands
OUTPut Subsystem
6.15 OUTPut Subsystem
The OUTPut subsystem controls the output features of the instrument.
In conjunction with the tracking generator option, a distinction is made between OUTPut1 (screen A) and
OUTPut2 (screen B).
OUTPut<1|2>[:STATe]
ON | OFF
This command switches the tracking generator on or off.
Example:
"OUTP ON"
'Switches on the tracking generator in screen A.
Characteristics: *RST value:
OFF
SCPI: conform
Mode:
R, A
This command is only valid in conjunction with option tracking generator R&S FSU-B9.
OUTPut:UPORt[:VALue]
#B00000000 to #B11111111
This command sets the control lines of the user ports. In manual operation, the control lines are
represented by softkeys PORT 0 to 7.
The user port is written to with the given binary pattern. If the user port is programmed to INPut instead
of OUTPut, the output value is temporarily stored.
Example:
"OUTP:UPOR #B10100101"
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
R, A
OUTPut:UPORt:STATe
ON | OFF
This command switches the control line of the user ports between INPut and OUTPut.
The user port is switched to OUTPut with parameter ON, to INPut with OFF.
Example:
"OUTP:UPOR:STAT ON"
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
R, A
• With the tracking generator switched on, the maximum stop frequency is limited
to 3 GHz. This upper limit is automatically modified by the set frequency offset
of the generator.
• If measurements in compliance with specs are to be performed with the tracking
generator, the start frequency has to be
≥
3 x resolution bandwidth.
• The minimum sweep time for measurements in compliance with the data sheet
is 100 ms in the frequency domain (span >0). If a shorter sweep time is
selected, the sweep time indicator SWT on the screen is marked with a red
asterisk and the message UNCAL is also displayed.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...