R&S ESU
Receiver Mode
4.28
Operating Manual 1302.6163.12 - 03
4.5.4.1
Selecting the Detector
Six different detectors can be selected for weighting the receive signal.
• The max peak detector yields the largest sample values of the levels measured
during the set measurement time.
• The min peak detector yields the smallest sample values of the levels measured
during the set measurement time.
• The average detector yields the average level of the samples measured during
the set measurement time.
• The CISPR average detector supplies a weighted average. When measuring the
average according to CISPR 16-1, the maximum value of the linear average
during the measurement time is displayed. The detector is used, for example, to
measure pulsed sinusoidal signals with a low pulse frequency. It is calibrated with
the rms value of an unmodulated sinusoidal signal. Averaging is with lowpass
filters of the 2nd order (simulation of a mechanical instrument). The lowpass time
constants and the IF bandwidths are fixed depending on the frequency. The main
parameters are listed in the following table:
Coupling of the IF bandwidth to the frequency range with the CISPR average
detector activated can be switched off by the
CISPR RBW UNCOUPLED
softkey.
• The rms detector yields the rms level of the samples measured. The integration
time corresponds to the set measurement time.
• The CISPR rms detector supplies a weighted average. When measuring the
average according to CISPR 16-1, the maximum value of the linear average
during the measurement time is displayed. The detector is used, for example, to
measure pulsed sinusoidal signals with a low pulse frequency. It is calibrated with
the rms value of an unmodulated sinusoidal signal. Averaging is with lowpass
filters of the 2nd order (simulation of a mechanical instrument). The lowpass time
constants and the IF bandwidths are fixed depending on the frequency. The main
parameters are listed in the following table:
The coupling of the IF bandwidth to the frequency range with activated CISPR
RMS detector can be cancelled using the
CISPR RBW UNCOUPLED
softkey.
Band A
Band B
Band C/D
Band F
Frequency range
< 150 kHz
150 kHz to
30 MHz
30 MHz to 1
GHz
> 1 GHz
IF bandwidth
200 Hz
9 kHz
120 kHz
1 MHz
Time constant of instrument
160 ms
160 ms
100 ms
100 ms
Band A
Band B
Band C/D
Band E
Frequency range
< 150 kHz
150 kHz to
30 MHz
30 MHz to 1
GHz
> 1 GHz
IF bandwidth
200 Hz
9 kHz
120 kHz
1 MHz
Time constant of instrument
160 ms
160 ms
100 ms
100 ms
Corner frequency
10 Hz
100 Hz
100 Hz
1 kHz
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...