Operating Manual 1302.6163.12 - 03
6.113
R&S ESU
Remote Control – Description of Commands
DISPlay Subsystem
DISPlay:BARGraph:LEVel:UPPer?
This command queries the maximum level of the bar graph.
Example:
":DISP:BARG:LEV:UPP?"
Characteristics:
*RST value: --
SCPI: device-specific
Mode:
R
DISPlay:BARGraph:PHOLd
ON | OFF
This command switches the indication of the maxhold value of the bar graph measurement on or off.
Example:
":DISP:BARG:PHOL ON"
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
R
DISPlay:BARGraph:PHOLd:RESet
This command resets the maxhold value of the numeric indication of the bar graph measurement.
Example:
":DISP:BARG:PHOL:RES"
Characteristics:
*RST value: --
SCPI: device-specific
Mode:
R
This command is an event and thus has no *RST value and no query.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...