R&S ESU
Remote Control – Description of Commands
DISPlay Subsystem
6.112
Operating Manual 1302.6163.12 - 03
Mode:
R, A
DISPlay[:WINDow<1|2>]:TRACe<1...3>:MODE:HCONtinuous
ON | OFF
This command specifies whether or not the traces with peak or minimum value detection are reset after
specific parameter changes.
Usually the measurement must be restarted after a parameter change, before an evaluation of the
measurement results is performed (e.g. with a marker). In cases in which a change causes a
compulsory new measurement, the trace is automatically reset in order to prevent erroneous
measurements of previous measurement results (e.g. when the span changes). For applications in
which this behavior is not desired, this mechanism can be switched off.
Parameter:
OFF: The traces are reset after specific parameter changes.
ON: The reset mechanism is switched off.
Example:
"DISP:WIND1:TRAC3:MODE:HCON ON"
'The reset mechanism is switched off for measurement window 1.
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A
DISPlay[:WINDow<1|2>]:TRACe<1...3>[:STATe]
ON | OFF
This command switches on or off the display of the corresponding trace in the selected measurement
window.
Example:
"DISP:WIND1:TRAC3 ON"
Characteristics:
*RST value: ON for TRACe1, OFF for TRACe2 and 3
SCPI: conform
Mode:
R, A
DISPlay[:WINDow<1|2>]:TRACe<1...3>:SYMBol
CROSs | OFF
This command switches on or off the indication of the peak list or the final measurement result on the
trace.
Parameter:
CROSs: The single value is indicated as x.
OFF: No indication.
Example:
"DISP:TRAC:SYMB CROS"
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
R
DISPlay:BARGraph:LEVel:LOWer?
This command queries the minimum level of the bar graph.
Example:
":DISP:BARG:LEV:LOW?"
Characteristics:
*RST value: --
SCPI: device-specific
Mode:
R
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...