Operating Manual 1302.6163.12 - 03
6.81
R&S ESU
Remote Control – Description of Commands
CALCulate Subsystem
CALCulate<1|2>:MARKer<1...4>:FUNCtion:SUMMary:MEAN:PHOLd:RESult?
This command queries the result of the measurement of the mean value with active peak hold in the
selected measurement window. The query is only possible if the peak hold function has been switched
on previously using
CALCulate<1|2>:MARKer<1...4>:FUNCtion:SUMMary:AVERage
The query is possible only if the peak hold function is active. The function is independent of the marker
selection, i.e. the numeric suffix <1...4> of
:MARKer
is irrelevant. It is only available in the time domain
(span = 0).
A complete sweep with synchronization to sweep end must be performed between switching on the
function and querying the measured value to obtain a valid query result. This is only possible in single-
sweep mode.
Example:
"INIT:CONT OFF"
'Switches to single-sweep mode
"CALC:MARK:FUNC:SUMM:MEAN ON"
'Switches on the function in screen A
"CALC:MARK:FUNC:SUMM:PHOL ON"
'Switches on the peak value measurement in screen A
"INIT;*WAI"
'Starts a sweep and waits for the end
"CALC:MARK:FUNC:SUMM:MEAN:PHOL:RES?"
'Outputs the result of screen A.
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
A-T
This command is only a query and therefore has no *RST value.
CALCulate<1|2>:MARKer<1...4>:FUNCtion:SUMMary:SDEViation[:STATe]
ON | OFF
This command switches on or off the measurement of the standard deviation in the selected
measurement window. The function is independent of the marker selection, i.e. the numeric suffix
<1...4> of
:MARKer
is irrelevant. It is only available in the time domain (span = 0).
On switching on the measurement, the mean power measurement is switched on as well.
Example:
"CALC2:MARK:FUNC:SUMM:SDEV ON"
'Switches on the measurement of the standard deviation in screen B.
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...