R&S ESU
Remote Control – Description of Commands
CALCulate Subsystem
6.40
Operating Manual 1302.6163.12 - 03
CALCulate<1|2>:LIMit<1...8>:UPPer:OFFSet
<numeric_value>
This command defines an offset for the y-axis of the selected relative upper limit line. In contrast to
CALC:LIM:UPP:SHIFt
, the line is not shifted by modifying the individual Y values but by means of
an additive offset. The offset is independent of the measurement window.
Example:
"CALC:LIM2:UPP:OFFS 3dB"
'Shifts limit line 2 by 3 dB upwards in the corresponding measurement windows.
Characteristics:
*RST value: 0
SCPI: device-specific
Mode:
R, A
CALCulate<1|2>:LIMit<1...8>:UPPer:MARGin
<numeric_value>
This command defines a margin to an upper limit line, at which out-of-limit values are signaled (if the
limit check is active), but not handled as a violation of the limit value. The margin is independent of the
measurement window.
Example:
"CALC:LIM2:UPP:MARG 10dB"
'Defines the margin of limit line 2 to 10 dB below the limit value.
Characteristics:
*RST value: 0
SCPI: device-specific
Mode:
R, A
CALCulate<1|2>:LIMit<1...8>:UPPer:MODE
RELative | ABSolute
This command selects the relative or absolute scaling for the y-axis of the selected upper limit line. The
setting is independent of the measurement window.
Selecting RELative causes the unit to be switched to DB.
Example:
"CALC:LIM2:UPP:MODE REL"
'Defines the y-axis of limit line 2 as relative scaled.
Characteristics:
*RST value: ABSolute
SCPI: device-specific
Mode:
R, A
CALCulate<1|2>:LIMit<1...8>:UPPer:SHIFt
<numeric_value>
This command moves a limit line by the indicated value in Y direction. In contrast to
CALC:LIM:UPP:
OFFS
, the line is shifted by modifying the individual Y values and not by means of an additive offset.
Up to 8 limit lines can be defined at the same time. The shift is independent of the measurement
window.
Example:
"CALC:LIM3:UPP:SHIF 20DB"
'Shifts all Y values of limit line 3 by 20 dB.
Characteristics:
*RST value: --
SCPI: device-specific
Mode:
R, A
This command is an event and therefore has no *RST value and no query.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...