R&S ESU
Instrument Functions
External Generator Control – Option R&S FSP-B10
4.338
Operating Manual 1302.6163.12 - 03
Fig. 4.32 Normalized measurement, shifted with REF VALUE POSITION 50%
Remote command:
DISP:WIND:TRAC:Y:RPOS 10PCT
REF VALUE
The
REF VALUE
softkey activates the input of a level difference which is assigned
to the reference line.
In the default setting, the reference line corresponds to a level difference of 0 dB. If
e.g. a 10-dB attenuator pad is inserted between DUT and R&S ESU input between
recording of the calibration data and normalization, the trace will be shifted down by
10 dB. By entering a
REF VALUE
of –10 dB the reference line for difference forma-
tion can also be shifted down by 10 dB so that it will again coincide with the trace
(see
).
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...