Operating Manual 1302.6163.12 - 03
4.295
R&S ESU
Instrument Functions
Basic Settings
EXPORT /
IMPORT DEV
DATA
The
EXPORT DEV DATA
softkey opens a dialog to export device specific data onto
USB stick or floppy disk.
The export function automatically creates a sub folder using the instrument name
and the serial number, e.g. FSQ40_123000 and copies device specific files to this
folder:
• Option license key codes
• Board specific data
• Frequency response correction file (option FSQ-B72 only)
IMPORT DEV DATA reads these data back onto the instrument.
Select the folder, where the instrument specific sub folder is located and press
OPEN
. A dialog asks to perform a reboot when the import is done.
This function is helpful to backup device specific data, e.g. the option license keys. It
is also helpful if you have to use a removable flash disk for more than one device. In
this case:
• Export the instrument data of device 1.
• Switch instrument 1 off.
• Remove the flash disk from device 1.
• Insert the flash disk into device 2.
• Switch instrument 2 on
• Import the instrument data of device 2 (saved before).
Note that a new export of the specific instrument data is required whenever:
• A new option key was inserted.
• The Rohde & Schwarz Service refitted a board.
• The Rohde & Schwarz Service performed a device calibration.
Remote command:
---
Selftest
SELFTEST
The
SELFTEST
softkey initiates the selftest of the instrument modules.
With this function the instrument is capable of identifying a defective module in case
of failure.
During the selftest a message box appears in which the current test and its result is
shown. The test sequence can be aborted by pressing ENTER ABORT.
All modules are checked consecutively and the test result (selftest PASSED or
FAILED) is output in the message box.
Remote command:
*TST?
SELFTEST
RESULTS
The
SELFTEST RESULTS
softkey calls the
SELFTEST
table in which the results of
the module test are displayed.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...