Operating Manual 1302.6163.12 - 03
4.193
R&S ESU
Instrument Functions
Analyzer Mode
As of firmware version 4.4x, the weigthing filter parameters can be manually con-
trolled for USER ACP measurements. To change the weighting filter settings when
using predefined standards, proceeed as follows:
➢
Select the predefined standard with the
softkey.
➢
Save the settings as a user defined standard with the
softkey.
➢
Select the user defined standard you just saved with the
softkey.
➢
Since the standard in use is now a user defined standard, you can change the
weighting filter settings as required.
➢
Save the user standard again to keep the current settings.
Weighting filters are not available for FAST ACP measurements and for Multi Carrier
ACP measurements.
WEIGHT TX
(ON OFF)
Activates and deactivates the weigthing filter for the TX channel.
Note that most of the standards require identical settings for all channels.
Remote command:
POW:ACH:FILT:STAT:ALL ON
POW:ACH:FILT:STAT:CHAN ON
WEIGHT TX
ALPHA
Defines the roll-off factor of the weighting filter for the TX channel.
Remote command:
POW:ACH:FILT:ALPH:ALL 0.3
POW:ACH:FILT:ALPH:CHAN 0.3
WEIGHT ADJ
(ON OFF)
Activates and deactivates the weigthing filter for adjacent and alternate channels
.Note that most of the standards require identical settings for all channels.
Remote command:
POW:ACH:FILT:STAT:ACH OFF
POW:ACH:FILT:STAT:ALT ON
WEIGHT ADJ
ALPHA
Defines the roll-off factor of the weighting filter for adjacent and alternate channels.
Remote command:
POW:ACH:FILT:ALPH:ACH 0.3
POW:ACH:FILT:ALPH:ALT 0.3
ADJ REFLVL
OFFSET
Defines an additional level offset to the measured mean power to be taken into
account for the analyzer's reference level setting.
It is a modification to the ADJUST REF LEVEL function specific to user standards.
Proceed as follows to adjust the offset value for a predefined standard:
➢
Select a predefine standard with the
Changing the TX channel filter parameter will additionally adjust the ADJ channel
settings as long as the current settings for TX and ADJ channels are identical. As
most of the digital standards require identical settings for all channels it is there-
fore required to configure the TX channel weigthing filter parameter only.
Change the ADJ weigthing filter settings first, if different settings are required for
TX and ADJ channels.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...