EDIT ACP
LIMITS
The
EDIT ACP LIMITS
softkey opens a table for defining the limits for the ACP mea-
surement.
The following rules apply for the limits:
• A separate limit can be defined for each adjacent channel. The limit applies to both
the upper and the lower adjacent channel.
• A relative and/or absolute limit can be defined. The check of both limit values can
be activated independently.
• The R&S ESU checks adherence to the limits irrespective of whether the limits are
absolute or relative or whether the measurement is carried out with absolute or
relative levels. If both limits are active and if the higher of both limit values is
exceeded, the measured value is marked accordingly.
Remote command:
CALC:LIM:ACP ON
CALC:LIM:ACP:ACH 0dB,0dB
CALC:LIM:ACP:ACH:STAT ON
CALC:LIM:ACP:ACH:ABS –10dBm,-10dBm
CALC:LIM:ACP:ACH:ABS:STAT ON
CALC:LIM:ACP:ALT1 0dB,0dB
CALC:LIM:ACP:ALT1:STAT ON
CALC:LIM:ACP:ALT1:ABS –10dBm,-10dBm
CALC:LIM:ACP:ALT1:ABS:STAT ON
CALC:LIM:ACP:ALT2 0dB,0dB
CALC:LIM:ACP:ALT2:STAT ON
CALC:LIM:ACP:ALT2:ABS –10dBm,-10dBm
CALC:LIM:ACP:ALT2:ABS:STAT ON
WEIGHTING
FILTER
Opens a submenu to control the weighting filter.
Some of the digital standards specify weighting filters to be taken into account. In
that case the required filter is automatically activated by the selection of the ACP
standard.
Measured values exceeding the limit are marked by a preceding asterisk.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...