ADJUST
SETTINGS
The
ADJUST SETTINGS
softkey automatically optimizes the instrument settings for
the selected power measurement (see below).
All instrument settings relevant for a power measurement within a specific frequency
range (channel bandwidth) are optimized for the selected channel configuration
(channel bandwidth, channel spacing):
• Frequency span:
The frequency span should cover at least all channels to be considered in a
measurement.
For channel power measurements, the frequency span is set as follows:
(No. of transmission channels - 1) × transmission channel s
2 × transmission channel ban measurement margin
For adjacent-channel power measurements, the frequency span is set as a
function of the number of transmission channels, the transmission channel
spacing, the adjacent-channel spacing, and the bandwidth of one of adjacent-
channels ADJ, ALT1 or ALT2, whichever is furthest away from the transmission
channels:
(No. of transmission channels - 1) × transmission channel s
2 × (adjacent-channel s adjacent-channel bandwidth) + measurement
margin
The measurement margin is approx. 10% of the value obtained by adding the
channel spacing and the channel bandwidth.
• Resolution bandwidth: RBW
≤
1/40 of channel bandwidth
• Video bandwidth: VBW
≥
3 × RBW
• Detector: RMS detector
Trace math and trace averaging functions are switched off.
The reference level is not influenced by
ADJUST SETTINGS
. It can be separately
adjusted with
ADJUST REF LVL.
The adjustment is carried out only once; if necessary, the instrument settings can be
changed later.
Remote command:
SENS:POW:ACH:PRES ACP|CPOW|MCAC|OBW
ACP LIMIT
CHECK
The
ACP LIMIT CHECK
softkey switches the limit check for the ACP measurement
on and off.
Remote command:
CALC:LIM:ACP ON
CALC:LIM:ACP:ACH:RES?
CALC:LIM:ACP:ALT:RES?
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...