Operating Manual 1302.6163.12 - 03
4.187
R&S ESU
Instrument Functions
Analyzer Mode
With the IBW method (
FAST ACP OFF),
the channel bandwidth limits are marked by
two vertical lines right and left of the channel center frequency. It can in this way be
visually checked whether the entire power of the signal under test is within the
selected channel bandwidth.
Measurements in the time domain (
FAST ACP ON)
are performed in the zero span
mode. The channel limits are indicated by vertical lines. For measurements requiring
channel bandwidths deviating from those defined in the selected standard the IBW
method is to be used.
Refer to section
for a list of available filters.
When measuring according to the IBW method (FAST ACP OFF) the bandwidths of
the different adjacent channels are to be entered numerically. Since all adjacent
channels often have the same bandwidth, the other channels Alt1 and Alt2 are set to
the bandwidth of the adjacent channel on entering the adjacent-channel bandwidth
(ADJ). Thus only one value needs to be entered in case of equal adjacent channel
bandwidths. The same holds true for the ALT2 channels (alternate channels 2) when
the bandwidth of the ALT1 channel (alternate channel 1) is entered.
Remote command:
SENS:POW:ACH:BWID:CHAN 14kHz
SENS:POW:ACH:BWID:ACH 1kHz
SENS:POW:ACH:BWID:ALT1 14kHz
SENS:POW:ACH:BWID:ALT2 14kHz
CHANNEL
SPACING
The
CHANNEL SPACING
softkey opens a table for defining the channel spacings
for the TX channels as well as for the adjacent channels.
TX channels
The spacing between every TX channels can be defined separately. Therefore a TX
spacing 1-2 for the spacing between the first and the second carrier, a TX spacing 2-
3 for the spacing between the second and the third carrier and so on can be defined.
In order to allow a convenient setup for the system with equal TX channel spacing,
the value of TX spacing 1-2 will be copied in all the spacing below after entry, the TX
spacing 2-3 will be copied in all the spacing below after entry and so forth.
The channel spacings can be set separately by overwriting the table from top to
bottom.
The entry "TX" is only available for the multi carrier ACP measurement.
The entry "TX" is only available for the multi carrier ACP measurement.
For different spacings, a setup from top to bottom is necessary.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...