R&S ESU
Instrument Functions
Analyzer Mode
4.172
Operating Manual 1302.6163.12 - 03
The number of sweeps over which the average or the maximum value is calculated
is set with the
NUMBER OF SWEEPS
softkey.
Remote command:
CALC:MARK:FUNC:SUMM:RMS ON
CALC:MARK:FUNC:SUMM:RMS:RES?
MEAN
The
MEAN
softkey switches on the calculation of the mean value from the points of
the displayed trace or a segment of it. The linear mean value of the equivalent volt-
ages is calculated.
This can be used for instance to measure the mean power during a GSM burst.
For the maximum peak, the largest mean value obtained since the activation of
MAX
HOLD ON
is displayed.
With
AVERAGE ON
, the mean values of a trace are averaged over several sweeps
and displayed.
The number of sweeps over which the average or the maximum value is calculated
is set with the
NUMBER OF SWEEPS
softkey.
Remote command:
CALC:MARK:FUNC:SUMM:MEAN ON
CALC:MARK:FUNC:SUMM:MEAN:RES?
STANDARD
DEVIATION
The
STANDARD DEVIATION
softkey switches on the calculation of the standard
deviation of trace points from the mean value and outputs them as measured value.
The measurement of the mean power is automatically switched on at the same time.
For the maximum peak, the largest standard deviation obtained since the activation
of
MAX HOLD ON
is displayed.
With
AVERAGE ON
, the standard deviations of a trace are averaged over several
sweeps and displayed.
The number of sweeps over which the average or the maximum value is calculated
is set with the
NUMBER OF SWEEPS
softkey.
Remote command:
CALC:MARK:FUNC:SUMM:SDEV ON
CALC:MARK:FUNC:SUMM:SDEV:RES?
LIMIT ON | OFF
The
LIMIT ON/OFF
softkey selects the limited (
ON
) or non-limited (
OFF
) evaluation
range.
The evaluation range is defined by the
START LIMIT
and
STOP LIMIT
softkeys. If
LIMIT = ON
, signals are only searched between the two lines.
If only one limit line is switched on, time line 1 is the lower limit and the upper limit
corresponds to the end of sweep. If time line 2 is also switched on, it defines the
upper limit.
If no limit line is switched on, the evaluation range is not limited.
The default setting is
LIMIT = OFF
.
Remote command:
CALC:MARK:X:SLIM OFF
START LIMIT
The
START LIMIT
softkey activates the entry of the lower limit of the evaluation
range.
Remote command:
CALC:MARK:X:SLIM:LEFT <value>
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...