Operating Manual 1302.6163.12 - 03
4.151
R&S ESU
Instrument Functions
Analyzer Mode
The delta marker1 x-value can not be changed away from 0 as long as the link func-
tionality is active.
Remote command:
CALC:DELT:LINK ON | OFF
CNT RESOL ...
The
CNT RESOL ...
softkeys select the counter resolution. They are selection
switches, i.e. only one of the can be active at any one time.
The marker stop time, i.e. the frequency measurement time, depends on the
selected resolution.
Remote command:
CALC:MARK1:COUN:RES <value>
Measurement example
The frequency of a CW signal is to be determined by means of the frequency
counter with a resolution of 10 Hz.
STEPSIZE
STANDARD
The
STEPSIZE STANDARD
softkey controls the knob increment/decrement of the
marker position and uses the grid resolution (span/625).
Remote command:
CALC:MARK:X:SSIZ STAN
STEPSIZE
SWP POINTS
The
STEPSIZE SWP POINTS
softkeys controls the knob increment/decrement of
the marker position and uses the available sweep points configured in the
SWEEP
menu.
Remote command:
CALC:MARK:X:SSIZ POIN
MKR FILE
EXPORT
The
MKR FILE EXPORT
softkey saves the data of all active markers of the window
to a specified file. The format of the decimal point is defined by the
DECIM SEP
soft-
key.
Remote command:
MMEM:STOR:MARK 'C:\marker.txt'
DECIM SEP
The
DECIM SEP
softkey selects the decimal separator between '.' (decimal point)
and ',' (comma) for the function
MKR FILE EXPORT
.
Remote command:
FORM:DEXP:DSEP POIN
[
PRESET
]
R&S ESU is set to the default setting.
[
MARKER
]
MARKER 1
is switched on and set to the maximum value of
the displayed spectrum.
[
SIGNAL COUNT
]
The frequency counter is switched on. R&S ESU counts the
frequency of the signal at the marker position with a resolu-
tion of 1 kHz. The counted frequency is indicated in the
marker info field.
[
NEXT
]
Changes to the submenu for setting the counter resolution.
[
CNT RESOL 10 Hz
]
The frequency counter resolution is increased to 10 Hz.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...