Operating Manual 1302.6163.12 - 03
4.125
R&S ESU
Instrument Functions
Analyzer Mode
Fig. 4.3 Pulsed signal GATE OFF
Fig. 4.4 TDMA signal with GATE ON
The gated-sweep mode is activated by the
GATED TRIGGER
softkey. The setting of
the mode takes place in the
GATE SETTINGS
submenu.
GATED
TRIGGER
The
GATED TRIGGER
softkey switches the sweep mode with gate on and off.
When gate is switched on, a gate signal applied to the rear panel connector
EXT
TRIGGER/GATE
or the internal IF power detector controls the sweep of the
R&S ESU. This selection is made via the
EXTERN
and
IF POWER
softkeys for trig-
ger and gate.