Operating Manual 1302.6163.12 - 03
4.123
R&S ESU
Instrument Functions
Analyzer Mode
FREE RUN
The
FREE RUN
softkey activates the free-run sweep mode, i.e. start of a sweep is
not triggered. Once a measurement is completed, another is started immediately.
FREE RUN
is the default setting of R&S ESU.
Remote command:
TRIG:SOUR IMM
VIDEO
The
VIDEO
softkey activates triggering through the displayed voltage.
For the video triggering mode, a horizontal trigger line is shown in the diagram. It
may be used to set the trigger threshold between 0% and 100% of the overall dia-
gram height.
Remote command:
TRIG:SOUR VID
TRIG:LEV:VID 50 PCT
EXTERN
The
EXTERN
softkey activates triggering via a TTL signal at the input connector
EXT TRIGGER/GATE
on the rear panel.
The external trigger level can be adjusted in the range from 0.5 V to 3.5 V. The
default value is 1.4 V.
Remote command:
TRIG:SOUR EXT <numeric_value>
SWE:EGAT:SOUR EXT
TRIG:SEQ:LEV:EXT
IF POWER
The
IF POWER
softkey activates triggering of the measurement via signals which
are outside the measurement channel.
For this purpose, the R&S ESU uses a level detector at the second intermediate fre-
quency. The detector threshold can be selected in a range between –30 dBm and -
10 dBm at the input mixer. The resulting trigger level at the RF input is calculated via
the following formula:
For this purpose, the R&S ESU uses a level detector at the second intermediate fre-
quency. The detector threshold can be selected in a range between –50 dBm and -
10 dBm at the input mixer. The resulting trigger level at the RF input is calculated via
the following formula:
The bandwidth at the intermediate frequency is 10 MHz. The R&S ESU is triggered
as soon as the trigger threshold is exceeded within a 5 MHz range around the
selected frequency (= start frequency in the frequency sweep).
Thus, the measurement of spurious emissions, e.g. for pulsed carriers, is possible
even when the carrier lies outside the selected frequency span.
Remote command:
TRIG:SOUR IFP
SWE:EGAT:SOUR IFP
Side menu
MixerLevel
min
RFAtt PreampGain
–
+
InputSignal MixerLevel
max
RFAtt PreampGain
–
+
≤
≤
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...