11
EN: AUTOMATIC ARM STYLE BLOOD PRESSURE MONITOR |
Instruction Manual
Note: The memory can not be cleared unless you cut the power off.
CARE AND MAINTENANCE
• Do not expose the BP monitor to extreme temperature, moisture, dust or direct
sunlight.
• The cuff contains a sensitive airtight bulb. Treat it with care and avoid any stress by
torsion or kinking.
• Clean the BP monitor with a soft cloth. Do not use fuel, thinner or other solvent.
Remove specks on the cuff carefully with a wet cloth and soap water
• Do not wash the cuff.
• Do not drop the BP monitor or treat it with force.
• Do not try to open the BP monitor . This will make the factory calibration invalid.
• If the unit has been stored at temperatures below 0°C(32°F), leave it in a warm place
for about 15 minutes before using it. Otherwise, the cuff may not inflate properly.
• Do not attempt to disassemble the unit.
• Do not crush the cuff.
• When the unit will not be used for a long time (30 days or more), be sure to remove
the batteries.
CALIBRATION
This product has been calibrated at the time of manufacture. If used according to the
use instructions, periodic recalibration is not required. If at any time, you question the
measuing accuracy, please contact an authorized Service Center.
ERROR MESSAGE
Message
Correction
The pressure measured was lower than 20mmHg. Please measure
again.
The pressure can not be measured due to signal noise. Please measure
again.
The pressure measured was higher than 300mmHg. Please measure
again.
Pumping error. Please check cuff and try again.
The pumping pressure is higher than 300mmHg. Please measure again.
or not
display
Low battery. Please check and replace the 4 batteries if necessary.
Er
1
Er
2
Er
3
Er
p
H1