Ramsey Medical petMAP+II Quick Reference Instructions Download Page 1

 
 

 

CONTROLS & CONNECTIONS  

 

 

 

OPERATING MODES & DISPLAYS 

  

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

USER SELECTIONS 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SD Card 
Slot 

 

ON/OFF 

 

Blood Pressure Measurement & Related Commands 

 

SpO2 & Related Commands 

 

 

 

 

OR MODE 

BP cycle set

 

 

CLINIC MODE 

 

DIRECTIONS FOR USE 

(abbreviated*)

 

1.

 

 Power the 

II

 :

 

Press and release the ON/OFF button on the right side of the unit to power 
the device on (See CONTROLS & CONNECTIONS section above). 

 

2.

 

Place the sensors.  

 

a.

 

BP cuff.  

Select the proper size cuff based on measurement site, animal’s 

size and cuff index markings.  Apply the cuff snuggly.  If necessary, attach 
the cuff to the proper extension, provided.   

b.

 

SpO2 sensor (if equipped). 

 Clip the sensor to the tongue or ear and 

touch the ON O2sat button to begin measuring.  

c.

 

Temperature (if equipped)

.  Insert the temperature probe orally into the 

esophagus or rectally.  The temperature value will be displayed on the 
right side of the screen just under the SpO2 heart rate (See OPERATING 
MODES & DISPLAYS section above).

  

      

  

  

3.

 

Set Up 

II

.

 

a.

 

Attach cuff, SpO2 cable and temperature connectors to the top of the 
unit (See CONTROLS & CONNECTIONS section above). 

b.

 

Press the Species/Cuff Site button repeatedly to change the different 
settings.  Stop when you arrive at the appropriate setting (See USER 
SELECTIONS section above). 

 

          

 

 

AC/DC Adapter  

Connection 

Quick Reference Instructions 

AC/DC Adapter 

4.

 

Measuring in the O.R.

   

a.

 

BP settings in OR Mode.

  BP cycle times can be set by touching the BPcycle 

button repeatedly until you arrive at the desired time interval. High and low 
systolic BP alarms can also be set.  To set alarms, repeatedly touch the BP 
HiAlarm and LoAlarm buttons until you reach the desired value (See Blood 
Pressure Measurement & Related Commands in USER SELECTIONS section 
above).  Press BP Go/Stop to initiate cycle.        

   

 

b.

 

SpO2 settings in OR Mode.  

A low alarm for SpO2 can be set by repeatedly 

touching the LoAlarm button in the second row of buttons next to the SpO2 
value.  Audible heart rate can also be set at 1 of 5 different volumes; touch 
the HrAudio button until you reach the desired volume (See SpO2 & Related 
Commands in the USER SELECTIONS section above).  

5.

 

Measuring in the Exam Room. 

 

For use in the exam room mode, leave the BP cycle setting OFF.  For each 
reading, touch the BP Go/Stop button.  In this mode a trend of readings, along 
with the NSV (after at least two readings), will appear on the bottom left of the 
screen and a history of the last 10 readings taken will be displayed on the 
bottom right of the screen.

        

*More detailed information is available in the Operator’s Manual.  A copy of the manual is available for download at 

www.petmap.com

   To contact Customer Service, call 813-289-5555; toll free 800-231-6370.   

SpO2 Connector 

Cuff Connector 

Temperature  
Connector 

Mini  

USB Port 

Charging indicator 

green – fully charged 
     amber - charging

 

 

Nominal Session 
Value (NSV) 

Systolic BP 

Diastolic BP 

Mean Arterial Pressure 

Heart rate (from BP) 

SpO2  & pleth 

SpO

Heart rate 

Temperature 

Systolic BP 

Diastolic BP 

Heart rate (from BP) 

SpO

Heart rate 

Measurement History 

Temperature 

Time of Day 

Battery status 

Species, Cuff site setting 

Cuff 
Pressure 

BP Trend

  

Temperature 

Trend 

Press and release to set the BP measurement frequency which, when set, places the device into OR mode in which BP reading are 
automatically repeated at the selected interval.  The available BPcycle options are STAT (BP cycles rapidly for three minutes then reverts 
automatically to a 2 min cycle time), 1 Min, 2 Min, 3 Min, 4 Min, 5 Min, 10 Min, 15 Min, 20 Min, 25 Min and 30 Min.  Once the desired 
frequency is selected, the user must press the BP Go/Stop button to begin the measurements.  Press and hold the BP cycle button to turn the 
cycle mode OFF.   

Press and release to set the low limit alarm for systolic pressure.  The available settings are 60, 80, 100, 120, 140, 160 and OFF. Press and hold to 
turn the alarm OFF. 

Press and release to set the high limit alarm for systolic pressure.  The available settings are 150, 175, 200, 225, 250 and OFF.  Press and hold to 
turn the alarm OFF.  

Press and release to set the species and cuff site which is displayed at the top.  The seven available settings are DOG (ForeArm, HindFoot and 
Tail), CAT (ForeArm, HindFoot and Tail) and Optimize…None.  The non-optimized mode should be used on any species other than dog or cat or 
when comparing petMAP to Doppler or other oscillometric BP devices. 

Press and release to initiate a BP reading in Clinic Mode, or to initiate a series of automatically cycled BP readings  
in the OR Mode. 
 

Press and release to start the measurement of O2 saturation.  When the SpO2 monitor is ON, the button will show ON and the plethysmographic 
waveform will be displayed, along with the SpO2 value and heart rate derived from the SpO2 sensor.  It may take 10-15 seconds for the display to 
stabilize after switching O2 ON. 

Press and release to set the low limit alarm for SpO2.  The available settings are 96, 94, 92, 90, 88, 86, 84, 82 and OFF. Press and hold to turn 
alarm OFF. 

Press and release to set the volume for the SpO2 audible heart rate cardiotachometer.  The available settings are ON 1 – ON 5 ( 5 being the 
loudest) and OFF.  Press and hold to turn the audio OFF. 

Time of Day

  

If an SD card is in use, press to capture an image of the screen in a file on the SD card.  Similarly, the unit will prompt the user at power off for the 
same feature.  (See Operator’s Manual) 
 
 