MTR 221 ACS - MTR 221 ACS DHHS
MANUALE ISTRUZIONI / INSTRUCTION MANUAL
S
YSTEM
S
TART
-U
P
58
Model R 221 ACS - R 221 ACS DHHS
MTR 221
ACS
-
MTR 221
ACS
DHHS
- Sy
stem Star
t-Up
Install
Firmware Upload Complete
Open the Loader_2.1.exe file and power ON
the collimator.
If the configuration stage has been correctly
completed, the upload procedure shall start
automatically.
When the X-Ray Collimator upload is
completed, the tool retrieves the proper status
message.
If the firmware uploading procedure is completed successfully, the X-Ray Collimator shall
reboot automatically. If not, reboot it.
Issues List
The Loader_2.1 application program can retrieve error messages in one of the following
scenarios:
1. Fail to open COM port.
Possible Root Causes of the problem:
• the seriale.txt file has not been properly configured.
• the USB Serial Converter is not correctly connected both to the PC and / or to the X-Ray
Collimator Serial Port.
2. Firmware Upload does not start.