Model R 221 ACS - R 221 ACS DHHS
MT
R
2
21
ACS
-
MTR
2
21
AC
S DHH
S - In
stal
lation
Alignment Device
This device is used to ensure the collimator
mounting flange is correctly aligned to the X-ray
beam. The X-ray tube manufacturer provides a
tolerance for the placement of X-ray tube focus.
Ralco recommends to use the Focal Alignment
Device (jig) to ensure the correct flange alignment
with the centre of the X-Ray beam, see
Fig. Focal
Alignment Device.
By making an exposure, it is
possible to verify the perpendicularity and
concentricity using fixed references on the X-Ray
image.
Once the mounting flange is aligned the collimator
light/X-ray field should also be aligned (within
specific tolerances).
Please consult the technical specifications of your X-
Ray tube to find the maximum tolerance for the
position of the focus. Should the use of an alignment
device not be possible, Ralco collimators allow for
the regulation of the light field.
Mounting the Collimator to the Flange
N
OTE
T
HE
MOUNTING
FLANGE
PROVIDED
(
IF
PURCHASED
)
WITH
THE
COLLIMATOR
IS
SUBJECTED
TO
TESTING
PURSUANT
TO
ALL
APPLICABLE
STANDARDS
.
M
OUNTING
B
RACKET
T
ABS
CONFORM
TO
EN60601.
W
ARNINGS
P
URSUANT
TO
APPLICABLE
STANDARDS
, R
ALCO
HAS
TESTED
THE
COLLIMATOR
AND
FLANGE
APPLYING
STATIC
LOADS
. R
ALCO
IS
NOT
IN
A
POSITION
TO
KNOW
THE
DYNAMIC
FORCES
OF
ALL
END
-
USER
SYSTEMS
. I
T
IS
THE
RESPONSIBILITY
OF
THE
END
-
USER
TO
ENSURE
DYNAMIC
FORCES
OF
THE
SYSTEM
DO
NOT
CREATE
A
DANGEROUS
CONDITION
.
A - X-Ray Tube, B - Focal Adjustment Device
C - Mounting Flange, D - Screw, E - Washer
Fig. Focal Alignment Device