Pipetman

®

-

F

single-volume microliter pipettes

Fixed volumes from
0.1 µL to 10 mL

F-200 shown

Summary of Contents for Pipetman F Series

Page 1: ...Pipetman F single volume microliter pipettes Fixed volumes from 0 1 L to 10 mL F 200 shown...

Page 2: ...ing Tip Ejector Arm 8 Replacement Parts 9 Parts Drawings 10 11 Performance Specifications 12 Service Calibration and Repair 12 Contacting Rainin Back Cover Technical Assistance 800 543 4030 Call this...

Page 3: ...for long trouble free service The body is made of high impact chemical resistant plastic with low ther mal conductivity Pipetman may be held for prolonged periods without affecting sample volume repro...

Page 4: ...a lint free tissue taking care not to touch the tip orifice 6 To dispense sample touch the tip end against the side wall of the receiving vessel and press the plunger slow ly to the FIRST STOP Wait 1...

Page 5: ...ically to ensure that every lot meets Rainin s high standards Precision Calibration Pipetman F can be calibrated in the laboratory for a spe cific solution viscosity or test This feature is very usefu...

Page 6: ...using an F 200 Pipetman to pipette a particular viscous solution you determine gravimetrically that the volume delivered is 196 5 L and the indicator setting reads F2 Therefore you wish to increase t...

Page 7: ...ility low variance This procedure is recommended when critical repro ducibility is required In short intertip and intratip vari ances will be minimized if pre rinsing is practiced Reverse Mode Pipetti...

Page 8: ...F 10ML use the special filters supplied Temperature Considerations You can measure warm or cold liquids with good preci sion by using a consistent pipetting rhythm A consistent rhythm will help to mi...

Page 9: ...n components with the exception of the grease seal on models F 5001 to 10ML prior to serial number N05803D Leaks Inaccuracy Abnormal Stroke Possible causes 1 Loosened shaft Tighten coupling by hand 2...

Page 10: ...http www rainin com pdf ab14 pdf Removing Replacing Tip Ejector Arm 2 Pull arm from handle using ejector arm release collar 1 Press tip ejector button Tip Ejector Arm Removal 1 Press tip ejector butt...

Page 11: ...positioner G Small spring 23871 23871 23871 23871 23871 positioner H Small spring 300066 300066 300047 300047 300047 J Seal assy 44817 44817 23354 44603 23306 holder K Polyethylene 44815 44818 23359 4...

Page 12: ...10 F 1MIC to F 2MIC F 2 1MIC to F 10MIC A E F G H J L K B C E F G H J K L A B C F 1 to F 20 F 21 to F 100 A E F G H J L K B C E F G H J K L A B C...

Page 13: ...11 F 1001 to F 5000 F 5001 to F 10ML B M M K J B J K F 101 to F 200 F 201 to F 1000 E F G H J K L A B C K L B C...

Page 14: ...5 F 400 3 60 1 20 0 80 0 20 F 500 4 00 0 80 1 00 0 20 F 1000 8 00 0 80 1 30 0 13 Specifications for models not listed are available from Technical Service 800 543 4030 Service Calibration and Repair R...

Page 15: ......

Page 16: ...anual Spec ified performance is guaranteed only when Rainin tips are used Contacting Rainin Technical Information Phone 800 543 4030 Fax 781 938 1152 E mail tech service rainin com Pipette Service Pho...