42
S_S
Select to set sample mass by determining mass of a single part.
Suu
Select to set sample mass by entering mass of a single part.
•
Enter respective value and press
key to confirm, then continue
weighing.
21.2. Setting Sample Mass by Entering Mass of a Single Part
•
Enter
<3.2.PcS / 3.2.2.UUt>
submenu, set
<
Suu
>
value.
•
Enter
<PcS>
working mode (parts counting), first, text
<SEt_Ut>
is displayed for 1 s, next, window for entering mass value of a single part.
•
Enter respective value and press
key to confirm, home screen
is displayed automatically along with quantity of parts loaded onto the
weighing pan (pcs).
If the value of entered single part mass is greater than max
capacity value, then message <Err Hi> is displayed.
21.3. Setting Sample Mass by Determining Mass of a Single Part
•
Enter
<3.2.PcS / 3.2.2.UUt>
submenu, set
<
S_S
>
value.
•
Enter
<PcS>
working mode (parts counting), blinking value of sample
quantity is displayed.
•
Press
key to select one of the following options:
10
Reference sample quantity: 10 pcs.
20
Reference sample quantity: 20 pcs.
50
Reference sample quantity: 50 pcs.
100
Reference sample quantity: 100 pcs.
0000
Custom reference sample quantity - enter the required value yourself.
•
Select respective option and press
key to confirm, first, text
<LoAd>
is displayed for 1 second, then the weighing window.
•
If the parts are to be weighed in a container, first put the container on
a weighing pan and tare it.