- 28 -
Procedure:
1. Enter <Adjustment> submenu, next select “
External adjustment” parameter.
2. “Remove weight” prompt is displayed.
3. Take the weight off the weighing pan and press
button. Whereas balance determines
start mass, “Start mass determination” message is displayed.
4. Upon completed start mass determination procedure “Put weight …” prompt is displayed along
with particular mass standard value.
5. Put the required weigh on a pan and press
button.
6. Upon completed procedure “Remove weight” prompt is displayed.
7. Take the weight off the weighing pan, press
button, the balance proceeds to the
weighing mode.
10.3. USER ADJUSTMENT
User adjustment is carried out with an optional standard of mass ranging between 0.15 Max and
Max. User adjustment and external adjustment procedures are likewise with one exception, before
user adjustment start a message box for entering mass of a standard used for user adjustment is
opened.
CAUTION!
User adjustment is possible for balances that are not a subject to conformity assessment (verification).
In order to start user adjustment, enter <Adjustment> submenu and select ‘
User adjustment’
parameter. Then follow the prompts displayed on a screen.
10.4. ADJUSTMENT TEST
<Adjustment test> function enables comparing the result of an internal automatic adjustment with the
value recorded in balance factory parameters. The comparison is used for determining balance
sensitivity drift over time.
10.5. AUTOMATIC ADJUSTMENT
This menu is for declaring a value initiating an automatic adjustment. Accessible options:
•
None – automatic adjustment disabled,
•
Time – adjustment takes place in time intervals declared in menu <Automatic adjustment time>
(10.6),
•
Temperature – adjustment is triggered by temperature change only,
•
Both – adjustment is triggered by both, temperature changes and time.
CAUTION!
Parameter settings can be modified only for balances that are not a subject to conformity assessment
(verification).
10.6. AUTOMATIC ADJUSTMENT TIME
<
Automatic adjustment time> parameter determines time interval between successive internal
adjustments carried out automatically. The time interval is declared in hours and ranges between 1
and 12 hours.
To set time interval for automatic adjustment:
•
select <Automatic adjustment time> parameter,