20
Options in case when the main unit is [kg]:
•
kg (kilogram),
•
lb (pound)*,
•
N
(newton)*,
•
u1 (custom unit 1)*,
•
u2 (custom unit 2)*.
*)
–
unit disabled for verified weighing devices.
You can declare start unit and two custom units – for detailed
information read section 15, „UNITS”.
8.5. Setting MIN, MAX Thresholds
MIN, MAX thresholds are used:
•
to control mass of the weighed loads (read section 20.4 of this user
manual),
•
for graphic visualization informing you how much of the weighing device
capacity is used (read section 13.1.3 of this user manual),
•
to control the external automation systems using digital outputs of the
indicator (read section 12.2 of this user manual).
The mass is considered to be correct if it is comprised within MIN and MAX
threshold values. You can set MIN, MAX thresholds as follows:
•
using
programmable
<
Set MIN, MAX>
button,
•
selecting product with set thresholds,
•
using a proximity sensor,
•
using a digital input.
MAX threshold value has to be greater than MIN threshold
value.
8.5.1. Setting MIN, MAX Threshold Using Programmable Button
•
Enter
<Display / Buttons functions>
submenu.
•
Edit a given button.
•
Select
<Set MIN and MAX>
parameter from the list.
•
Exit to the home screen.
•
Press previously programmed button,
<Min>
numerical box is opened.
Summary of Contents for PUE HX5.EX