18
8. WEIGHING
Load the weighing pan. Read the result when
stability marker is
displayed.
Only stable weighing results can be recorded (stability marker
).
8.1. Zeroing
To zero mass indication press
key. Zero indication and the following
pictograms are displayed:
and
.
Zeroing operation means determining new zero point, recognized by the
weighing device as precise zero. The instrument can be zeroed only when the
indication is stable.
Indication can be zeroed only within ±2% range of maximum
capacity. If the zeroed value is greater than ±2% of the
maximum capacity, then the software displays a respective
error message:
8.2. Taring
To determine net weight value, load the weighing pan with a packaging, wait
for a stable indication and press
key. Zero indication and the following
pictograms are displayed:
and
. The weighing device has been
tared.
Remember not to exceed scale's maximum capacity, i.e. sum of tare weight
value and load weight value must be lower than the maximum capacity value.
Upon unloading the weighing pan, the sum of tared masses with minus sign
is displayed.
You can assign tare value to a particular product stored in the database.
In such case the assigned tare value is automatically acquired upon selection
of the given product.