8.3.
Connecting Load Cell’s Cable Shield
Load cell with galvanic
connection of the signal
cable shield
Load cell without galvanic
connection of the signal
cable shield
Scale with an indicator in a plastic
housing - the platform connected with the
indicator via load cell's signal cable only.
do not connect
E
Scale of compact design (scale with an
indicator in a plastic housing on a post)
do not connect
E
Scale with an indicator in a metal housing
- the platform connected with the indicator
via load cell's signal cable only.
Point A
Point A
E
– solder pad on the main board and additional A/D converter boards
.
Point A
– threaded pin, connected with indicator's housing galvanically (metal housing).
9. OPTIONAL EXTENSION MODULES
9.1. Additional Platform Module, DP6
The DP6 module expands indicator functionality, with it an additional platform
can be operated. The module is intended to be mounted inside the indicator.
Along with the DP6 module (484R board), in the indicator's housing cover an
additional grommet is installed, through it the additional platform cable is fed.
9.1.1. Technical Specifications
Operating temperature
-
10 °C ÷ 40 °C
OIML
III
Verification unit [e]
6000
Maximum input signal
19.5 mV
Maximum voltage per verification unit
3.25 µV
Minimum voltage per verification unit
0.4 µV
Minimum load cell impedance
50 Ω
Maximum load cell impedance
1200 Ω
Load cell power supply
5 V
Load cell wiring
4 or 6 wires + Shield
Multi range option
Yes